General Lab

pH Meter Calibration: Best Practices for Accuracy

DDr. Evelyn Reed
July 05, 2024
pH Meter Calibration: Best Practices for Accuracy

Accurate pH measurement is critical in many pharmaceutical processes, from buffer preparation to final product formulation. This article outlines the best practices for calibrating your pH meter.

1. Use Fresh, Certified Buffers

The accuracy of your calibration is directly tied to the accuracy of your buffers. Always use fresh, certified pH buffers (e.g., pH 4, 7, and 10). Avoid using buffers that are expired or have been exposed to air for long periods, as their pH can shift due to CO2 absorption.

2. The Calibration Process

For most applications, a two-point calibration is sufficient. Always bracket your expected sample pH. For example, if you expect a pH of 8.5, calibrate with pH 7 and pH 10 buffers. Rinse the electrode thoroughly with deionized water and blot dry (do not wipe!) between each buffer. Allow the reading to stabilize in each buffer before confirming the calibration point.

3. Checking Slope and Offset

After calibration, check the electrode's slope and offset values. The slope should be between 95% and 105%, and the offset should be within ±30 mV. Values outside this range may indicate an aging or dirty electrode that needs cleaning or replacement.

Test Your Knowledge

Click the button below to generate a quiz based on this article using AI.